分流/不分流进样口
分流/不分流进样口
要查看 7890/6890/6850 分流/不分流进样口备件的图解零部件明细表,请 单击此处
对于食品检测、环境和毒理学实验室中从事痕量分析工作的分析人员,可将我们专有的气相沉积 UltiMetal Plus* 应用于惰性焊件。如果要增加进样量来进行痕量分析,那么对流路中的热金属表面进行脱活处理很重要,因为样品将溢出衬管,分析物将暴露于金属表面中。
*请注意,脱活处理产生的颜色差异是正常的。
- 进样口
产品详细信息
- 应用简报
-
Evaluating Inert Flow Path Components and Entire Flow Path for GC/MS/MS Pesticide Analysis
Flow path inertness is critical in pesticide analysis accuracy and precision, especially for active analytes such as organophosphate pesticides.
- 应用简报
- English
- 2013 年 9 月 12 日
- 1.46 MB
Trace analysis of volatile organic acids with the Agilent J&W DB-624UI GC column
Agilent’s new deactivation of the 6% cyanopropyl dimethylpolysiloxane (624) phase significantly improves acid performance and maintains very good performance.
- 应用简报
- English
- 2013 年 8 月 9 日
- 2.43 MB
Agilent J&W DB-624UI 超高惰性色谱柱对挥发性有机酸的痕量分析
安捷伦的新型去活 6% 氰丙基二甲基聚硅氧烷 (624) 固定相极大地提高了对酸性物质的分离性能,而且保留了对碱性物质和醇类物质的出色分离性能
- 应用简报
- Chinese (Simplified)
- 2013 年 3 月 27 日
- 1.66 MB
- 产品样本
- 数据表
- 宣传单页
- 挂图
- 参考数据
- 技术概述
- 药品说明书
-
Parts Finder Insert (Combined Version English, Japanese, Chinese)
Find your ideal match of Agilent GC Columns and Supplies with the Agilent Parts Finder tool and the GC Column Selection Tool.
- 药品说明书
- English
- 2013 年 6 月 27 日
- 2.72 MB
Find your ideal match: GC Tools promotion insert
Find your ideal match of Agilent GC Columns and Supplies with the Agilent Parts Finder tool and the GC Column Selection Tool.
- 药品说明书
- English
- 2013 年 6 月 27 日
- 1.12 MB
惰性流路全览
- 00:02:48
- 2013 年 3 月 11 日
To help you get the most out of your Agilent products.
www.agilent.com/chem/agilentresources
Split Inlet Troubleshooting
Split inlets are spared from most band-broadening phenomena, since the splitting process generates narrow peaks. Peak broadening or tailing is usually due to:
• Improper column installation
• Low inlet temperature
• Low split flow (<20 mL/min on 6890)
• Inlet and needle discrimination and decomposition
如果您的结果不准确或者不一致:
• Check the column and reinstall if necessary
• Increase inlet temperature by 50°C and compare results
• Check inlets and needles for wear and replace as necessary
不分流进样口故障排除
Most problems encountered with a splitless injection are related to:
• Incorrect purge time
• Degradation
• Improper focusing
• Inappropriate column temperature
• Backflash
You can also improve the reproducibility and linearity of peak areas and avoid backflash by matching:
• Inlet temperature
• Liner volume
• Injection volume